The Telaire T6700 Series consists of miniature NDIR CO2 Sensors that have the accuracy and reliability of many larger sensors. The new small size allows OEM’s to integrate in to smaller enclosures and equipment and uses significantly less power than many other devices on the market. All units are factory calibrated to measure CO2 concentration levels up to 5000ppm, while maintaining accuracy across the range.
Within the T6700 Series:
-
Telaire T6703 CO2 Module is configured for applications where CO2 levels are less critical but an assessment of indoor air quality is required, for example in residential applications. Our minimum order requirement reflects the high volumes of these applications.
-
Telaire T6713 CO2 Module is ideal for applications where accurate CO2 levels need to be measured and controlled for indoor air quality and energy saving applications, such as demand control ventilation.
- Telaire T6723-8K5 CO2 Alarm Module is designed for heater applications requiring high limit controls.
Applications
Ideal for use where CO2 levels need to be measured and controlled for indoor air quality and energy saving applications, such as demand control ventilation.
T6703 Applications: |
T6713 Applications: |
T6723-8K5 Applications: |
|
|
|
Features
- An affordable gas sensing solution for OEMs.
- Eliminates the need for calibration in most applications with Telaire’s patented ABC LogicTM software. Lifetime calibration warranty.
- A reliable sensor design based on 20 years of engineering and manufacturing expertise.
- Flexible CO2 sensor platform designed to interact with other microprocessor devices.
- Compact design and versatile interface options for simple product integration.
Specs T6703
Method | Non Dispersive Infrared (NDIR), gold plated optics,diffusion sampling (with Telaire’s Patented ABC Logic Self Calibrated Algorithm) |
---|---|
Measurement Range | 0 to 5000 ppm |
Dimensions | 1.18 in X 0.61 in X 0.34 in (30 mm X 15.6 mm X 8.6 mm) |
Accuracy* | 400-5000 ppm ± 75 ppm or 10% of reading, whichever is greater |
Temperature Dependence | 5 ppm per °C or 0.5% of the reading per °C, whichever is greater |
Stability | < 2% of FS over life of sensor (15 years typical) |
Pressure Dependence | 0.13% of reading per mm Hg |
Calibration Interval | Not required |
Response Time | < 3 minutes for 90% step change typical |
Signal Update | Every 5 seconds |
Warm Up Time |
|
Operating Conditions |
|
Storage Conditions | -22°F to 158°F (-30°C to 70°C) |
Specs T6713
Method | Non Dispersive Infrared (NDIR), gold plated optics, diffusion sampling (with Telaire’s Patented ABC Logic Self Calibrated Algorithm) |
---|---|
Measurement Range | 0 to 5000 ppm |
Dimensions | 1.18 in X 0.61 in X 0.34 in (30 mm X 15.6 mm X 8.6 mm) |
Accuracy* | 400-5000 ppm ± 30 ppm + 3% of reading |
Temperature Dependence | 5 ppm per °C or 0.5% of the reading per °C,whichever is greater |
Stability | < 2% of FS over life of sensor (15 years typical) |
Pressure Dependence | 0.13% of reading per mm Hg |
Calibration Interval | Not required |
Response Time | < 3 minutes for 90% step change typical |
Signal Update | Every 5 seconds |
Warm Up Time |
|
Operating Conditions |
|
Storage Conditions | -22°F to 158°F (-30°C to 70°C) |
Specs T6723-8K5
General Specifications: | |
---|---|
Operating Temperature Range | -20°C to +55°C |
Storage Temperature Range | -20°C to +70°C |
Operating Humidity Range | 0-99% |
Sensor Life | Designed for 5 years of heater use with no calibration required |
Self-Diagnostics | Detector and lamp failures |
"Forced Alarm" Test Pad | Simulates warning and alarm shutdown through the sensor |
Electrical Specifications: | |
---|---|
Power Input | 4.3VDC – 7.0VDC |
Average Current @ 6.0 VDC |
|
Peak Current | < 100mA ± 10% |
Low Voltage Shutdown | 4.3 VDC ± 5% |
Connector Type | JST B5B-EH-A(LF) |